Line


     A second exposure of the same metal samples reported in Testing Series B was performed over a longer period of 42 days. The purpose again was to measure the relative corrosion rate of the most commonly used CorrView ® models in 1-1/2 in. and 3/4 in. ASTM 1018 steel, and commercially available ASTM 1010 and 1018 corrosion coupons typically used in coupon racks to estimate pipe wall losses.


     As previously defined, a salt spray test booth was constructed. Design of the test booth substantially followed ASTM Designation B 117 - 95: Standard Practice for Operating Salt Spray (Fog) Apparatus.

     Coupons of ASTM 1010 and 1018 mild carbon steel, typically used in the corrosion rate evaluation of HVAC piping systems, were purchased from Metal Samples. Samples of actual ASTM A53, A106, and A795 pipe were also acquired on the open market, and fabricated into the approximate length and width of the corrosion coupons. Weights of the individual samples were not recorded, since an ultrasonic measurement of true wall loss would be used as the basis for corrosion rate, rather than weight loss.


Side View Of Test Assembly

     The wall thickness of one end of each corrosion coupon and pipe sample coupon was measured ultrasonically at 15 individual locations according to a standard grid pattern of specific dimension. Both CorrView ® models were similarly measured according to a specified grid at 15 individual locations at its center. All wall thickness data was recorded into a spreadsheet to establish a baseline.

     A glass enclosure was constructed to house the metal test specimens. All specimens were positioned on a wooden base for isolation and placed into the fog chamber. Specimens were physically arranged at an angle to the fog spray according to the requirements of ASTM B 117 - 95.


     A 5% salt spray solution was introduced into the test chamber via ultrasonic atomization and an overhead header configured to provide uniform and indirect introduction of the salt fog. Our corrosion simulation procedure took exception to maintaining a test chamber temperature of 95 ° F or to maintain pH, as the purpose of the test was to make a side by side comparison of different metal types rather than produce a standardized corrosion rate environment. All test subjects were exposed to the same conditions during its entire test period.

     Testing was continued and the results observed. On the first cycle of exposure at 42 days, the samples were removed and allowed to dry, then photographed. Deposits from the measured end of each test sample were then brushed free of rust deposits and rinsed. Following the removal of this area of deposits, ultrasonic testing was performed at the same area of each sample, and along the same grid pattern to provide a second set of wall thickness measurements. Wall loss and corrosion rates were calculated for each test specimen. Additional photographs were taken of the cleaned test samples.


     Results showed a general agreement of corrosion rate statistics within what we would consider reasonable limits. The salt spray environment produced an extremely aggressive attack at all metal surfaces as evidenced by visual inspection of the exposed metals, and of their underlying surfaces once deposits were removed (shown below). Moderate to high pitting was found in all examples, requiring the use of "echo to echo" ultrasonic measurement technique in order to accurately measure the base dimension of such areas.

     Corrosion rates of between 31.2 MPY and 61.8 MPY were measured; higher than those identified in Testing Series B and Testing Series C. The highly corrosive environment produced by the salt spray chamber would be expected to exaggerate minor differences in metal chemistry of the steel samples, and therefore produce a wider variation in result than under more typical conditions of a 1-3 MPY corrosion rate found at an open cooling tower system. The differences noted between the steel pipe samples alone help support this position.

     We would have expected lower corrosion rates for this longer sample exposure given that virgin steel corrodes at an aggressive rate for its first initial few days, and then declines to a more constant rate after some period of time. However, we can also raise the possibility that variances in temperature of the test setup between both series of tests, and possibly other different physical conditions, may have contributed to this difference. More important than differences between series of tests, however, is that the set of samples comprising each test set were exposed to identical conditions and therefore offer a comparative view of their respective corrosion susceptibility.

     Nevertheless, this series of tests showed a maximum variance of only 36.5% from the average corrosion rate measured, which we found reasonable. This highest variance in result was again found at sample # 5 of ASTM A53 pipe. The average variance of all samples tested from the mean corrosion rate was 22.5% - minor in comparison to the large discrepancy often found between corrosion coupons and true corrosion losses.

     Test results for the standard corrosion coupons of ASTM 1010 and ASTM 1018 mild steels showed similar rates of corrosion to the actual pipe samples measured. Installed in externally located corrosion coupon racks, however, their reported corrosion rates are typically far below what actually exists at the interior pipe wall - the result of their being isolated from most of the corrosive effects existing within the actual piping system.

     A corrosion rate determination for a condenser water system of 0.5 MPY using coupons, when the actual measured loss of pipe wall is 0.105 in. over 12 years or 8.75 MPY, is a substantial 1,750 % difference or under reporting of actual wall loss. Such extreme error in corrosion coupon rates, often found to exist between 100% and 1,000% in actual UT testing comparisons, shows the current 36.5 % variance in corrosion rate results of all samples tested to be nearly insignificant.

     The below graph well illustrates the variation in wall loss for the nine test subjects evaluated under this exposure period.




     The below set of tables offers a visual and statistical comparison of corrosion rates for the two corrosion coupons, four pipe sample coupons, and two CorrView ® products tested. Shown at the left is the original metal sample prior to testing, and in its original form. The center photograph is the exposed and rusted sample as removed from the salt fog booth. The far right photograph shows the same exposed sample after wire brushing.

     Our testing produced an average corrosion rate of 43.3 MPY at the five samples of actual steel pipe - ASTM A106, A795, A135, and A53, and an average corrosion rate of all nine metals tested of 43.3 MPY. For purposes of this evaluation, we then compared the corrosion rates of the commercially available steel corrosion coupons and CorrView ® products, finding an average 20.4 % and 9.8 % variance in their corrosion rates from the true pipe samples respectively. Both the corrosion coupons and CorrView products slightly over reported the corrosion rates found at the steel samples during this test series - opposite of what was found in Testing Series B.

     The percentage of variation in corrosion rate from the average measured value of 43.3 MPY measured at the five steel pipe samples is also provided below. Further testing is in progress, and those additional results will be presented when available.



  Metal Sample # 1 - Corrosion Coupon 1010 Steel


Specimen - Standard corrosion coupon from Metal Samples. ASTM 1010 mild carbon steel.
Exposure - 42 Days
Original Wall Thickness - 0.066 in.
Wall Loss - 0.0099 in. at two surfaces / Corrosion Rate - 42.9 MPY
Percent Variation From Mean Pipe Corrosion Rate - 6.3 %



  Metal Sample # 2 - Corrosion Coupon 1018 Steel


Specimen - Standard corrosion coupon from Metal Samples. ASTM 1018 mild carbon steel.
Exposure - 42 Days
Original Wall Thickness - 0.063 in.
Wall Loss - 0.0081 in. at two surfaces / Corrosion Rate - 35.1 MPY
Percent Variation From Mean Pipe Corrosion Rate - 23.4 %



  Metal Sample # 3 - Steel Pipe ASTM A106


Specimen - 2 in. ASTM A106 Grade B mild carbon steel pipe, schedule 10.
Exposure - 42 Days
Original Wall Thickness - 0.100 in.
Wall Loss - 0.0093 in. at two surfaces / Corrosion Rate - 40.3 MPY
Percent Variation From Mean Pipe Corrosion Rate - 12.0 %



  Metal Sample # 4 - Steel Pipe ASTM A795


Specimen - 1-1/2 in. ASTM A795 Grade A mild carbon steel pipe, schedule 10.
Exposure - 42 Days
Original Wall Thickness - 0.104 in.
Wall Loss - 0.0113 in. at two surfaces / Corrosion Rate - 49.2 MPY
Percent Variation From Mean Pipe Corrosion Rate - 7.6 %



  Metal Sample # 5 - Steel Pipe ASTM A53


Specimen - 1-1/4 in. ASTM A53 Grade B mild carbon steel pipe, schedule 40.
Exposure - 42 Days
Original Wall Thickness - 0.130 in.
Wall Loss - 0.0142 in. at two surfaces / Corrosion Rate - 61.8 MPY
Percent Variation From Mean Pipe Corrosion Rate - 35.1 %



  Metal Sample # 6 - Steel Pipe ASTM A53


Specimen - 2 in. ASTM A53 Grade B mild carbon steel pipe, schedule 40.
Exposure - 42 Days
Original Wall Thickness - 0.151 in.
Wall Loss - 0.0073 in. at two surfaces / Corrosion Rate - 31.7 MPY
Percent Variation From Mean Pipe Corrosion Rate - 30.7 %



  Metal Sample # 7 - 1-1/2 in. NPT CorrView Monitor


Specimen - CorrView ® Model # P3S1-NF1-BC, 1-1/2 in. NPT, 1018 mild carbon steel
Exposure - 42 Days
Original Wall Thickness - 0.099 in.
Wall Loss - 0.0068 in. at one surface / Corrosion Rate - 59.1 MPY
Percent Variation From Mean Pipe Corrosion Rate - 29.1 %



  Metal Sample # 8 - 3/4 in. NPT CorrView Monitor


Specimen - CorrView ® Model # P1S2-NF1-BC, 3/4 in. NPT, 1018 mild carbon steel.
Exposure - 42 Days
Original Wall Thickness - 0.052 in.
Wall Loss - 0.0012 in. at one surface / Corrosion Rate - 29.2 MPY
Percent Variation From Mean Pipe Corrosion Rate - 29.1 %


     Additional corrosion rate comparison testing is presented on this Internet site under the above Testing heading, and is currently underway. A duplicate of this testing procedure has been commissioned to an independent laboratory. Results will be presented here as soon as it is available. A third set of results from this group of test samples will be available on approximately 10/1/04.


©  Copyright