| |
|
 |
|
 |
|
|
|

|


Interest to
establish the corrosion characteristics of CorrView under well
maintained chemical treatment conditions prompted further testing of the same
type metal specimens evaluated under Series B, C, D, and F. Testing involved
CorrView ® models in 1-1/2 in. and 3/4 in. ASTM 1018
steel, and commercially available ASTM 1010 and 1018 corrosion coupons
typically used in coupon racks to estimate pipe wall losses.
In addition, comparison was
simultaneously made against the most commonly installed pipe for HVAC and
process water systems - ASTM A53, A106, A135, and A795. Testing was not
performed using any recognized corrosion testing method.

A two gallon
clear plastic container was used to house the 9 test samples. Coupons of ASTM
1010 and 1018 mild carbon steel, typically used in the corrosion rate
evaluation of HVAC piping systems, were purchased from Metal Samples.
Examples of actual ASTM A53, A106, A135,
and A795 pipe were also acquired on the open market, and fabricated into sample
coupons. Weights of the individual samples were not recorded, since an
ultrasonic measurement of true wall loss would be used as the basis for
corrosion rate, rather than weight loss.
 |
|
Above View Of Test
Assembly |
The wall
thickness of one end of each corrosion coupon and pipe sample coupon was
measured ultrasonically at 15 individual locations according to a standard grid
pattern of specific dimension. Both CorrView ® models
were similarly measured according to a specified grid at 15 individual
locations at its center. All wall thickness data was recorded into a
spreadsheet to establish a baseline.
A
small hole was drilled into one end of each pipe coupon in order to provide a
means to suspend the sample into the test chamber. Monofilament line was
attached to each test sample and marked to indicate its origin. Test samples #
1-7, comprising actual corrosion coupons and pipe materials were suspended into
the chamber with sufficient isolation between samples. The CorrView
® samples were rested on the bottom of the test chamber with
their wear surfaces facing upward.
A small
circulating pump was introduced into the upper area of the test chamber in
order to provide a continuous movement of water against all samples.

Chemical
corrosion inhibitor "Sureguard 4500" was provided by
Sentry Water
Management Corporation of Wayne, NJ. This formulation provided an
approximate 200 PPM concentration of sodium molybdate and 6-10 PPM of sodium
triazole. The plastic container was filled to ensure complete coverage of all
test subjects. Neither the temperature nor pH of the solution was maintained at
any specific value, and instead allowed to equalize to ambient conditions. All
test subjects were exposed to the same conditions during the entire test
period.
Testing was continued and the
results observed. On the first cycle of exposure at 15 days, the samples were
removed from the test chamber and allowed to dry, then photographed. No rust
deposits were observed on any metal samples and no wall loss was evident.

Results showed no
wall loss for any sample of metal over the 15 day exposure period. Ultrasonic
thickness testing of the test subjects showed no measurable wall loss and
therefore a zero corrosion rate has been assumed for
each.
While it is possible that some small
degree of wall loss may have occurred at some samples, and may be measurable
using a weight loss analysis rather than wall thickness measurement, any such
loss would be extremely small. A measurement of wall loss would likely be
possible over a more extended period of time, and for that reason, this testing
is being extended over an additional 90 day
period.
Our testing shows dramatic
difference in results from prior salt fog tests, as would be expected. While
the salt fog method is intended as an accelerated test for corrosion
susceptability, and produces far greater wall losses than under normal
corrosive conditions, the difference between Series G tests and those shown in
Series B, C, D, and F well illustrate the benefit of good corrosion control
methods.


The below set of
tables offers a visual and statistical comparison of results for the two
corrosion coupons, five pipe sample coupons, and two CorrView
® products tested. Shown at the left is the original metal
sample prior to testing, and in its original form. The right side photograph
shows the same sample after exposure, and without any surface preparation or
modification to its surface.
Our 15 day
exposure shows no measureable amount of wall loss at any of the test subjects.
We show zero wall loss for all samples under well maintained chemical inhibitor
conditions. A visual comparison of the samples shows minor discoloration for
some of the test subjects, but no rust product.
Metal Sample # 1 -
Corrosion Coupon 1010 Steel
 |
 |
Specimen - Standard corrosion coupon
from Metal Samples. ASTM 1010 mild carbon steel. Exposure - 15
Days Original Wall Thickness - 0.065 in. Wall Loss - 0.000 in. at two
surfaces Corrosion Rate - 0.0 MPY Percent
Variation From Mean Pipe Corrosion Rate - 0% |

Metal Sample # 2 -
Corrosion Coupon 1018 Steel
 |
|
Specimen - Standard corrosion coupon
from Metal Samples. ASTM 1018 mild carbon steel. Exposure - 15
Days Original Wall Thickness - 0.063 in. Wall Loss - 0.000 in. at two
surfaces Corrosion Rate - 0.0 MPY Percent
Variation From Mean Pipe Corrosion Rate - 0% |

Metal Sample # 3 -
Steel Pipe ASTM A106
 |
|
Specimen - 2 in. ASTM A106 Grade B
mild carbon steel pipe, schedule 10. Exposure - 15 Days Original Wall
Thickness - 0.099 in. Wall Loss - 0.000 in. at two surfaces Corrosion Rate - 0.0 MPY Percent Variation From Mean
Pipe Corrosion Rate - 0% |

Metal Sample # 4 -
Steel Pipe ASTM A795
 |
|
Specimen - 1-1/2 in. ASTM A795 Grade
A mild carbon steel pipe, schedule 10. Exposure - 15 Days Original Wall
Thickness - 0.104 in. Wall Loss - 0.000 in. at two surfaces Corrosion Rate - 0.0 MPY Percent Variation From Mean
Pipe Corrosion Rate - 0% |

Metal Sample # 5 -
Steel Pipe ASTM A53
 |
|
Specimen - 1-1/4 in. ASTM A53 Grade B
mild carbon steel pipe, schedule 40. Exposure - 15 Days Original Wall
Thickness - 0.131 in. Wall Loss - 0.000 in. at two surfacesCorrosion Rate - 0.0 MPY Percent Variation From Mean
Pipe Corrosion Rate - 0% |

Metal Sample # 6 -
Steel Pipe ASTM A53
 |
|
Specimen - 2 in. ASTM A53 Grade B
mild carbon steel pipe, schedule 40. Exposure - 15 Days Original Wall
Thickness - 0.149 in. Wall Loss - 0.000 in. at two surfaces Corrosion Rate - 0.0 MPY Percent Variation From Mean
Pipe Corrosion Rate - 0% |

Metal Sample # 7 -
Steel Pipe ASTM A135
 |
|
Specimen - 2-1/2 in. ASTM A135 Grade
B mild carbon steel pipe, schedule 10. Exposure - 15 Days Original Wall
Thickness - 0.108 in. Wall Loss - 0.000 in. at two surfaces Corrosion Rate - 0.0 MPY Percent Variation From Mean
Pipe Corrosion Rate - 0% |

Metal Sample # 8 -
1-1/2 in. NPT CorrView Monitor
 |
 |
|
Specimen - CorrView
® Model # P3S1-NF1-BC, 1-1/2 in. NPT, 1018 mild carbon
steel Front Wear Thickness - 0.098 in. Exposure - 15 Days Wall Loss -
0.0041 in. at one surface Corrosion Rate - 0.0
MPY Percent Variation From Mean Pipe Corrosion Rate -
0% |

Metal Sample # 9 -
3/4 in. NPT CorrView Monitor
 |
 |
|
Specimen - CorrView
® Model # P1S2-NF1-BC, 3/4 in. NPT, 1018 mild carbon
steel Front Wear Thickness - 0.051 in. Exposure - 15 Days Wall Loss -
0.000 in. at one surface Corrosion Rate - 0.0
MPY Percent Variation From Mean Pipe Corrosion Rate -
0% |
Additional
corrosion rate comparison testing is presented on this Internet site under the
above Testing
heading, and is currently underway. Results will be presented here as soon as
it is available. A second set of results from this group of test samples will
be available on approximately 05/1/05.
©
Copyright |
|
|
|
|
|
|

|
|
|
 |
|
|